.General Introduction

Yokogawa PH450 pH analyzer is a 4 wire multivariable analyzer combines pH with temperature and ORP measurement with advanced temperature compensation, and all these measurements can be utilized through the different output functions: two mA current outputs, four independent SPDT contact outputs and HART®. Both DD and DTM files are available for direct connection to HART® Handheld terminal, HMI monitor and Pact ware PC configuration.

 01                                      

2.General Specifications

Items

Description

Measuring principle

Electrochemical

Input ranges

pH: -2 to 16 pH
  ORP: -1500 to 1500 mV
  rH: 0 to 100rH
  Temperature: -30 to 140 deg C (Pt1000)

Transmission signals

General: Two isolated   outputs of 4 to 20mA DC with common negative. Maximum load 600 Ω.   Bi-directional HART® digital communication, superimposed on mA1 (4 to 20mA)   signal.
  Output function: Linear or Non-linear (21-step table) output for pH,   temperature, ORP or rH.
  Control function: PID control.
  Burn out function: Burn up (21.0mA) or burn down (3.6mA) to signal failure   accorded with NAMUR NE43.
  Hold: The mA-outputs are frozen to the last/fixed value during   calibration/commissioning.

Contact outputs

General: Four SPDT   relay contacts with display indicators.
  Switch capacity: Maximum values 100 VA, 250 V AC, 5 A; Maximum values 50   W, 250 V DC, 5 A
  Status: High/Low process alarms, selected from pH, ORP, rH and temperature.   Configurable delay time and hysteresis. Failure annunciation.
  Control function: On/Off, PID duty cycle or pulsed frequency control.
  Wash: Contact can be used to start manual- or interval time wash cycles.
  Hold: Contact can be used to signal the Hold situation.
  Fail: Contact S4 is programmed as fail-safe contact.

Contact input

Remote wash cycle   start.

Temperature   compensation Function

Automatic or manual.   Compensation to Nernst equation. Process compensation by configurable   temperature coefficient, NEN6411 for water or strong acids/bases or   programmable matrix.

Calibration

Semi-automatic 1 or 2   point calibration using pre-configured NIST, US, DIN buffer tables 4, 7 &   9, or with user defined buffer tables, with automatic stability check. Manual   adjustment to grab sample.

Display

Graphical Quarter VGA   (320 x 240 pixels) LCD with LED backlight and touch screen.

Housing

Cast Aluminum housing   with chemically resistant coating; Polycarbonate cover with Polycarbonate   flexible window, Protection IP66/NEMA 4X/CSA Type 3S

Power supply

PH450G-A:
    Ratings; 100 to 240 V AC, Acceptable range; 90 to 264 V AC
    Ratings; 50/60 Hz, Acceptable range; 50 Hz ±5%, 60 Hz ±5 %
    Power Consumption; 15 VA
  PH450G-D:
    Ratings; 12 to 24 V DC, Acceptable range; 10.8 to 26.4 V DC
    Power Consumption; 10 W

Environment and   operational conditions

Ambient temperature:   -20 to 55℃ (-5 to 130℉)
  Storage temperature: -30 to 70℃ (-20 to 160℉)
  Humidity: 10 to 90 % RH at 40℃ (100℉) (non-condensing)


3. Model Selection

Main Code

Suffix Codes

Description

PH450G




pH / ORP Converter

Power Supply

-A



AC version (100…240 VAC)

-D



DC version (12…24 VDC)

Type


-A


General purpose version


-U


FM version

Options



/UM

Universal mounting kit(panel, pipe, wall)



/U

Pipe and wall mounting hardware



/PM

Panel mounting hardware



/H5

Awning hood (stainless steel)



/AFTG

G1/2



/ANSI

1/2NPT



/AM20

M20



/SCT

Stainless steel tag plate

 

4.External Dimensions

02